脊髓手术D波导航:从监测到决策

李鹃 ,  奚健 ,  蒋星军

中南大学学报(医学版) ›› 2026, Vol. 51 ›› Issue (2) : 326 -332.

PDF (542KB)
中南大学学报(医学版) ›› 2026, Vol. 51 ›› Issue (2) : 326 -332. DOI: 10.11817/j.issn.1672-7347.2026.250622
综述

脊髓手术D波导航:从监测到决策

作者信息 +

D-wave-guided navigation in spinal cord surgery: From monitoring to decision-making

Author information +
文章历史 +
PDF (554K)

摘要

脊髓手术的核心目标是在最大安全切除病变部位的同时保护神经功能。直接波(direct wave,D波)监测作为评估皮质脊髓束(corticospinal tract,CST)功能完整性的金标准,已从被动监测工具发展为术中主动功能导航系统。为系统梳理D波导航功能的演进与应用,从以下3个方面进行阐述:技术革新(如硬膜下记录)保障了信号的稳定性与定量性;多模态整合(特别是与运动诱发电位协同)实现了突触前与突触后损伤的鉴别诊断;应用范围则已成功拓展至低龄儿童及多种复杂脊髓手术,突破了传统的年龄和病种限制。进一步探讨人工智能预警、影像融合导航等前沿方向,展望D波监测从经验依赖向数据驱动的智能化发展前景,可为提升脊髓手术的精准性与安全性提供理论支撑与实践指引。

Abstract

The primary goal of spinal cord surgery is to achieve maximal safe resection of lesions while preserving neurological function. Direct wave (D-wave) monitoring, regarded as the gold standard for assessing the functional integrity of the corticospinal tract (CST), has evolved from a passive monitoring tool into an active intraoperative functional navigation system. To systematically review the evolution and clinical application of D-wave-guided navigation, this article discusses 3 key aspects. First, technological advancements, such as subdural recording techniques, have improved signal stability and enabled quantitative assessment. Second, multimodal integration, particularly in combination with motor evoked potentials, facilitates the differential diagnosis between presynaptic and postsynaptic injuries. Third, the scope of application has been successfully expanded to include very young children and a wide range of complex spinal cord surgeries, overcoming traditional limitations related to age and disease type. Furthermore, emerging directions such as artificial intelligence-based warning systems and image-guided fusion navigation are explored, highlighting the transition of D-wave monitoring from experience-based practice to data-driven intelligence. These developments may provide both theoretical support and practical guidance for improving the precision and safety of spinal cord surgery.

关键词

直接波 / 术中监测 / 功能导航 / 皮质脊髓束 / 多模态整合 / 人工智能

Key words

direct wave / intraoperative monitoring / functional navigation / corticospinal tract / multimodal integration / artificial intelligence

引用本文

引用格式 ▾
李鹃,奚健,蒋星军. 脊髓手术D波导航:从监测到决策[J]. 中南大学学报(医学版), 2026, 51(2): 326-332 DOI:10.11817/j.issn.1672-7347.2026.250622

登录浏览全文

4963

注册一个新账户 忘记密码

脊髓手术因其功能结构的致密性与不可再生性,被公认为神经外科最具挑战性的领域之一[1-2]。现代手术理念强调在彻底切除病变部位的同时,最大程度保留神经功能并保障患者术后生活质量[3]。术中神经电生理监测(intraoperative neurophysiological monitoring,IONM)技术是实现此目标的关键保障[4]。其中,经颅电刺激运动诱发电位(transcranial motor evoked potentials,tcMEP)所记录的肌肉运动诱发电位(muscle motor evoked potentials,mMEP)虽应用广泛,但易受麻醉等因素干扰,对长期运动功能预后的预测价值有限[5]。相比之下,直接波(direct wave,D波)是经颅电刺激后于脊髓记录的皮质脊髓束(corticospinal tract,CST)快传导轴突的复合动作电位,其信号稳定可靠,振幅与功能完整的CST轴突数量呈线性正相关,被誉为评估CST结构完整性、预测术后长期运动功能的“金标准”[6]。一项于2025年发表、纳入1 079名患者的荟萃分析[7]进一步证实,D波在预测髓内占位永久性运动功能障碍方面的诊断效能显著优于mMEP。也有研究[8]定量地证实了术中D波变化与术后运动功能评分存在显著相关性,强化了其作为预后生物标志物的地位。随着技术进步与策略完善,D波监测已从单一的“损伤警报器”演进为全面的“功能导航系统”。本文旨在系统梳理D波导航在技术演进、多模态整合与应用拓展3大维度的进展,并展望其与前沿技术融合的未来方向。

1 D波功能导航的基石:技术演进与生理学基础

1.1 功能导航的生理学基石:稳定性与可定量性

D波通过“突触前机制”直接沿轴突下传,规避了麻醉药物的抑制[6, 9]。这赋予了其卓越的稳定性(术中变异度通常低于10%)和客观的定量评估能力(振幅与轴突数量线性相关)[1, 10]。术中D波振幅较基线下降50%,在理论上提示约半数功能完好的CST轴突可能受损,为术者量化评估CST损伤提供了独一无二的客观标尺[11],是实现精准功能导航的物理基础。

1.2 记录技术的革命性进展:从硬膜外到硬膜下

传统硬膜外记录常因高阻抗、脂肪或瘢痕组织导致信号衰减、信噪比低下[12]。近年来,硬膜下电极记录技术取得突破。研究[6]显示,其可将D波记录成功率从不足30%大幅提升至80%以上。将电极置于硬膜下腔能显著降低阻抗、提高基线振幅和信噪比,极大提升了信号可靠性[13],为在复杂病例中实施精准功能导航奠定了坚实基础。硬膜外与硬膜下记录的D波参数对比见表1。

1.3 预警标准的精细化:从单一阈值到分级管理

基于D波振幅与轴突数量的线性关系,国际公认的报警阈值为振幅较基线下降超过50%[14]。现代监测策略更趋向于分级预警,例如将下降30%~<50%、50%~70%、>70%分别定义为轻、中、重度预警[15]。这种精细化管理的核心价值在于提供了宝贵的“干预时间窗”。一项具有里程碑意义的历史对照研究[16]显示,采用分级预警后,术者在出现轻度预警时即采取干预措施,使患者术后永久性神经功能缺损的发生率从8%降至3%以下。一项回顾性研究[17]也支持这一分级策略的有效性,实现了导航功能从“损伤报警”到“风险预警”的关键前移。

2 功能导航的决策核心:多模态整合与智能解读

2.1 D波与mMEP的协同导航逻辑:突触前与突触后的鉴别诊断

D波的最大价值在于与mMEP等多模式构成协同监测体系[18]。二者联合解读的核心在于对神经损伤进行“突触前”与“突触后”的精准定位与鉴别诊断(表2)。当术中出现mMEP消失但D波保留(振幅>50%基线)时,D波的存在赋予了术者继续谨慎切除的“导航信心”,患者发生永久性运动功能障碍的风险很低(<2%)[19]。反之,当D波出现显著衰减(<50%基线)时,无论mMEP状态如何,均提示CST正遭受结构性损伤的高风险,是强烈的“终止导航”信号[19-20]。最新的荟萃分析[7]量化了这一价值:对于预测永久性运动功能障碍,D波的诊断效能显著高于mMEP(对数诊断比值比3.98 vs 1.89),其高特异度(95.0% vs 67.6%)是关键。但D波灵敏度低于mMEP(64.2% vs 73.8%),高特异度意味着当D波报警时,出现永久性损伤的可能性极高(假阳性率低),这对“终止手术”的决策至关重要。一项大型伞状综述[21]综合分析多项证据后指出,IONM(包括D波)总体上是安全有效的,能够为手术决策提供关键信息。

2.2 针对不同预后模式的精准导航

IONM变化组合能精准预测功能障碍的转归(表2)[7, 22]。若术后即刻功能障碍伴有显著D波变化,其发展为永久性的可能性极高(约87.5%);若仅伴mMEP变化而D波保留,则更可能是一过性的(恢复率约74.1%)[7]。这凸显了D波在预测长期转归方面的决定性作用[23]。需注意的是,二者各有侧重:mMEP在检测上肢精细运动及单侧CST损伤方面更敏感[7];而D波作为总体传导功能的指标,对单侧损伤可能不敏感[24]。因此,综合解读方能实现全面导航[25-26]。

2.3 人工智能赋能导航决策:从经验依赖向数据驱动演进

传统的IONM信号解读高度依赖个人经验,存在主观性和延迟性。人工智能(artificial intelligence,AI),尤其是深度学习技术,为D波导航的智能化带来了革命性前景[27]。基于循环神经网络(recurrent neural network,RNN)或长短期记忆网络(long short-term memory,LSTM)的模型能对D波信号进行毫秒级实时分析,综合评估波形形态、潜伏期微细变化及频谱特征,捕捉人眼难以识别的早期损伤模式[28]。已有研究[29]利用机器学习挖掘出比单一振幅下降阈值更精准的预后预测因子。对IONM数据进行全面、综合的解读,是发挥其最大效用的基础,这一原则同样适用于AI模型的构建[30]。未来,整合多模态IONM数据的AI决策支持系统,有望推动D波导航全面迈向数据驱动的新阶段[31]。

3 功能导航疆域拓展:超越脊髓髓内肿瘤的普适性验证

3.1 挑战极限:低龄儿童脊髓手术中的精准导航

传统观点[32]认为,低龄儿童因CST髓鞘化不完全,难以记录到可靠D波。通过个体化技术适配(如调整刺激电极位置、适度增加刺激强度、优化滤波设置),这一界限已被成功突破。Fekete等[33]成功为1例10月龄的脊髓栓系患儿实施了稳定的D波监测,术后患儿运动功能完好。近期针对儿童及青少年群体的研究[34]进一步优化了监测策略,并证实了D波在该人群中应用的可行性与价值。这表明,技术参数设置合适时,D波同样能为神经系统发育尚未成熟的儿科患者提供客观、可靠的功能导航,有效弥补了mMEP在该群体中的局限性[35]。

3.2 超越髓内:其他复杂脊髓手术中的宏观导航价值

D波监测的价值已超越经典的脊髓髓内肿瘤(intramedullary spinal cord tumor,IMSCT)切除术,在各类复杂手术中得到验证与拓展。最新的中国专家共识[36]也强调了包括D波在内的多模态IONM对于脊髓胶质瘤等疾病手术安全性的重要意义:1)髓外硬膜内肿瘤切除术。对于大型或腹侧肿瘤,手术牵拉是主要风险。D波振幅变化可客观反映脊髓对牵拉的耐受度,为术者判断牵拉极限提供量化参考,辅助平衡肿瘤切除与功能保护[37-38]。2)脊髓空洞症引流与脊髓栓系松解术。在此类减压手术中,D波可实时评估脊髓切开、粘连松解等操作对CST功能的影响,避免“减压性损伤”,提高手术安全性[39]。3)复杂脊柱畸形矫形手术。在严重侧凸或后凸矫形中,D波对脊髓整体传导功能的宏观、连续评估具有独特价值。D波可以作为mMEP和体感诱发电位的重要补充,尤其在监测因缓慢牵拉或血供变化导致的渐进性脊髓功能障碍方面可能更为敏感[40-41],为防止过度矫正提供又一道安全防线。

4 未来展望:迈向智能化与融合化的下一代功能导航系统

4.1 影像-电生理融合导航:从“听见”功能到 “看见”结构

将术前弥散张量成像(diffusion tensor imaging,DTI)纤维束示踪技术显示的CST三维走行信息,与术中实时获取的D波功能信息进行融合,是未来发展的重要方向[42]。这种“结构-功能”一体化导航系统,有望实现解剖与功能双重引导下的精准操作。

4.2 大数据与精准预后预测

利用机器学习等算法,分析海量术中D波数据(如振幅下降斜率、恢复模式)与患者长期功能结局之间的关系,有望构建精细化的预后预测模型,推动脊髓手术向精准医疗发展[29, 43]。

4.3 技术普及与标准化:导航作用广泛惠及的基石

尽管优势显著,D波监测在国内的普及度和规范化程度仍有待提高[44]。推动该技术的规范化与普及化,形成适合中国国情的专家共识与操作规范[17, 35, 45],并加强人员培训与标准化流程建设,是确保其导航价值能够稳定、可靠地惠及更多患者的关键[17]。在此过程中,参考和借鉴国际权威组织发布的最新临床实践指南至关重要[46-48]。

5 结 语

D波监测已从一项先进的电生理监测技术,蜕变为脊髓手术中不可或缺的“功能导航系统”。其基于神经生理学的稳定性与定量性、硬膜下记录等技术革新带来的可靠性提升、在多模态整合中展现出的损伤定位与预后预测智慧,以及在多年龄段、多类型脊髓手术中不断拓展的应用疆域,共同奠定了其不可动摇的地位[7, 19, 30]。未来,通过与AI、影像导航等前沿技术的深度融合,D波监测必将引领脊髓外科进入一个更精准、更安全、更智能的新时代[49]。

参考文献

[1]

Deletis V, Kothbauer K. Intraoperative neurophysiology of the corticospinal tract[M]//Spinal cord monitoring. Vienna: Springer Vienna, 1998: 421-444.

[2]

Samartzis D, Gillis CC, Shih P, et al. Intramedullary spinal cord tumors: part I: epidemiology, pathophysiology, and diagnosis[J]. Glob Spine J, 2015, 5(5): 425-435.

[3]

Hussain I, Parker WE, Barzilai O, et al. Surgical management of intramedullary spinal cord tumors[J]. Neurosurg Clin N Am, 2020, 31(2): 237-249.

[4]

Hadley MN, Shank CD, Rozzelle CJ, et al. Guidelines for the use of electrophysiological monitoring for surgery of the human spinal column and spinal cord[J]. Neurosurgery, 2017, 81(5): 713-732.

[5]

MacDonald DB, Skinner S, Shils J, et al. Intraoperative motor evoked potential monitorin-a position statement by the American Society of Neurophysiological Monitoring[J]. Clin Neurophysiol, 2013, 124(12): 2291-2316.

[6]

Deletis V, Sala F. Intraoperative neurophysiological monitoring of the spinal cord during spinal cord and spine surgery: a review focus on the corticospinal tracts[J]. Clin Neurophysiol, 2008, 119(2): 248-264.

[7]

Thomas S, Addanki RND, Murugan CK, et al. A systematic review and meta-analysis to compare the diagnostic accuracy of direct-waves and myogenic motor-evoked potential neuromonitoring during intramedullary spinal cord tumor resection[J]. Oper Neurosurg, 2026, 30(4): 513-524.

[8]

Quiñones-Hinojosa A, Lyon R, Zada G, et al. Changes in transcranial motor evoked potentials during intramedullary spinal cord tumor resection correlate with postoperative motor function[J]. Neurosurgery, 2005, 56(5): 982-993.

[9]

Legatt AD, Emerson RG, Epstein CM, et al. ACNS guideline: transcranial electrical stimulation motor evoked potential monitoring[J]. J Clin Neurophysiol, 2016, 33(1): 42-50.

[10]

Kothbauer KF. Intraoperative neurophysiologic monitoring for intramedullary spinal-cord tumor surgery[J]. Neurophysiol Clin, 2007, 37(6): 407-414.

[11]

Deletis V. Intraoperative monitoring of the functional integrity of the motor pathways[J]. Adv Neurol, 1993, 63: 201-214.

[12]

Costa P, Peretta P, Faccani G. Relevance of intraoperative D wave in spine and spinal cord surgeries[J]. Eur Spine J, 2013, 22(4): 840-848.

[13]

Ghadirpour R, Nasi D, Iaccarino C, et al. Intraoperative neurophysiological monitoring for intradural extramedullary spinal tumors: predictive value and relevance of D-wave amplitude on surgical outcome during a 10-year experience[J]. J Neurosurg Spine, 2018, 30(2): 259-267.

[14]

Morota N, Deletis V, Constantini S, et al. The role of motor evoked potentials during surgery for intramedullary spinal cord tumors[J]. Neurosurgery, 1997, 41(6): 1327-1336.

[15]

Olmsted ZT, Ryu B, Phayal G, et al. Direct wave intraoperative neuromonitoring for spinal tumor resection: a focused review[J]. World Neurosurg X, 2022, 17: 100139.

[16]

Sala F, Palandri G, Basso E, et al. Motor evoked potential monitoring improves outcome after surgery for intramedullary spinal cord tumors: a historical control study[J]. Neurosurgery, 2006, 58(6): 1129-1143.

[17]

中国医师协会神经外科医师分会神经电生理学组. 脊髓脊柱手术中神经电生理监测专家共识(2022版)[J]. 中华神经外科杂志, 2022, 38(4): 329-335.

[18]

Neuroelectrophysiology Group, Branch of Neurosurgeons, Chinese Medical Doctor Association. Expert consensus on neuroelectrophysiological monitoring in spinal and spinal cord surgery (2022 edition)[J]. Chinese Journal of Neurosurgery, 2022, 38(4): 329-335.

[19]

Stecker MM. A review of intraoperative monitoring for spinal surgery[J/OL]. Surg Neurol Int, 2012, 3():S174-S187[2025-11-01].

[20]

Kothbauer KF, Deletis V, Epstein FJ. Motor-evoked potential monitoring for intramedullary spinal cord tumor surgery: correlation of clinical and neurophysiological data in a series of 100 consecutive procedures[J/OL]. Neurosurg Focus, 1998, 4(5): E3[2025-11-01].

[21]

Cannizzaro D, Mancarella C, Nasi D, et al. Intramedullary spinal cord tumors: the value of intraoperative neurophysiological monitoring in a series of 57 cases from two Italian centers[J]. J Neurosurg Sci, 2022, 66(5): 447-455.

[22]

Rezapour A, Derakhshani N, Pouyan SN, et al. Safety and efficacy of intraoperative neuromonitoring: an umbrella review[J/OL]. Health Sci Rep, 2025, 8(10): e71370[2025-11-01].

[23]

Kimchi G, Knoller N, Korn A, et al. Delayed variations in the diagnostic accuracy of intraoperative neuromonitoring in the resection of intramedullary spinal cord tumors[J/OL]. Neurosurg Focus, 2021, 50(5): E21[2025-11-01].

[24]

Ushirozako H, Yoshida G, Imagama S, et al. Efficacy of transcranial motor evoked potential monitoring during intra- and extramedullary spinal cord tumor surgery: a prospective multicenter study of the monitoring committee of the Japanese society for spine surgery and related research[J]. Global Spine J, 2023, 13(4): 961-969.

[25]

Siller S, Duell S, Tonn JC, et al. Multimodal intraoperative neurophysiological monitoring may better predict postoperative distal upper extremities’ complex-functional outcome than spinal and muscular motor evoked potentials alone in high-cervical intramedullary spinal cord tumor surgery[J]. Clin Neurophysiol, 2024, 168: 52-60.

[26]

Li K, Tao X, Liu J, et al. The application value and improved warning criterion of D-wave monitoring in intramedullary spinal cord tumor surgery[J]. Spine J, 2024, 24(9): 1704-1711.

[27]

Velayutham P, Rajshekhar V, Chacko AG, et al. Influence of tumor location and other variables on predictive value of intraoperative myogenic motor-evoked potentials in spinal cord tumor surgery[J]. World Neurosurg, 2016, 92: 264-272.

[28]

Park D, Kim I. Application of machine learning in the field of intraoperative neurophysiological monitoring: a narrative review[J]. Appl Sci, 2022, 12(15): 7943.

[29]

Krieg SM, Shiban E, Droese D, et al. Predictive value and safety of intraoperative neurophysiological monitoring with motor evoked potentials in glioma surgery[J]. Neurosurgery, 2012, 70(5): 1060-1071.

[30]

Parduzi Q, Wermelinger J, Koller SD, et al. Explainable AI for intraoperative motor-evoked potential muscle classification in neurosurgery: bicentric retrospective study[J/OL]. J Med Internet Res, 2025, 27: e63937[2025-11-01].

[31]

Galloway GM, Nuwer MR, Lopez JR, et al. Intraoperative Neurophysiologic Monitoring[M/OL]. Cambridge: Cambridge University Press, 2010[2025-11-01].

[32]

Danilov GV, Shifrin MA, Kotik KV, et al. Artificial intelligence in neurosurgery: a systematic review using topic modeling. part I: major research areas[J]. Sovrem Tehnol Med, 2020, 12(5): 106.

[33]

Sala F, Krzan MJ, Deletis V. Intraoperative neurophysiological monitoring in pediatric neurosurgery: why, when, how?[J]. Childs Nerv Syst, 2002, 18(6/7): 264-287.

[34]

Fekete G, Bognár L, Novák L. D-wave recording during the surgery of a 10-month-old child[J]. Childs Nerv Syst, 2014, 30(12): 2135-2138.

[35]

Seidel K, Deletis V, Raabe A, et al. Intraoperative neurophysiologic monitoring and mapping during surgery on intramedullary spinal cord tumors in children and adolescents[J]. J Clin Neurophysiol, 2024, 41(2): 116-122.

[36]

Antkowiak L, Putz M, Sordyl R, et al. Relevance of intraoperative motor evoked potentials and D-wave monitoring for the resection of intramedullary spinal cord tumors in children[J]. Neurosurg Rev, 2022, 45(4): 2723-2731.

[37]

中国抗癌协会神经肿瘤专业委员会脊髓脊柱肿瘤学组, 中国医师协会神经外科医师分会脊髓脊柱专业组, 世界华人神经外科协会脊柱脊髓专业委员会. 脊髓胶质瘤诊疗中国专家共识[J]. 中华医学杂志, 2024, 104(37): 3478-3489.

[38]

Spinal and Spinal Cord Tumor Group of the Committee of Neuro‑Oncology of China Anti‑Cancer Association, Spinal Cord Specialty Group of the Branch of Neurosurgeons of Chinese Medical Doctor Association, Spine and Spinal Cord Committee of the World Chinese Neurosurgical Association. Chinese expert consensus on diagnosis and treatment of spinal cord glioma[J]. National Medical Journal of China, 2024, 104(37): 3478-3489.

[39]

Kumar N, Vijayaraghavan G, Ravikumar N, et al. Intraoperative neuromonitoring (IONM): is there a role in metastatic spine tumor surgery?[J/OL]. Spine, 2019, 44(4): E219-E224[2025-11-01].

[40]

D’Ercole M, D’Alessandris QG, Di Domenico M, et al. Is there a role for intraoperative neuromonitoring in intradural extramedullary spine tumors? results and indications from an institutional series[J]. J Pers Med, 2023, 13(7): 1103.

[41]

Paradiso G, Lee GY, Sarjeant R, et al. Multimodality intraoperative neurophysiologic monitoring findings during surgery for adult tethered cord syndrome: analysis of a series of 44 patients with long-term follow-up[J]. Spine, 2006, 31(18): 2095-2102.

[42]

Fehlings MG, Brodke DS, Norvell DC, et al. The evidence for intraoperative neurophysiological monitoring in spine surgery: does it make a difference?[J/OL]. Spine, 2010, 35: S37-S46[2025-11-01].

[43]

Ulkatan S, Neuwirth M, Bitan F, et al. Monitoring of scoliosis surgery with epidurally recorded motor evoked potentials (D wave) revealed false results[J]. Clin Neurophysiol, 2006, 117(9): 2093-2101.

[44]

Potgieser AR, Wagemakers M, van Hulzen AL, et al. The role of diffusion tensor imaging in brain tumor surgery: a review of the literature[J]. Clin Neurol Neurosurg, 2014, 124: 51-58.

[45]

Obermeyer Z, Emanuel EJ. Predicting the future: big data, machine learning, and clinical medicine[J]. N Engl J Med, 2016, 375(13): 1216-1219.

[46]

中国医师协会神经外科分会神经电生理监测专家委员会. 中国神经外科术中电生理监测规范(2017版)[J]. 中华医学杂志, 2018, 98(17): 1283-1293.

[47]

Expert Committee on Intraoperative Neurophysiological Monitoring, Branch of Neurosurgeons, Chinese Medical Doctor Association. Chinese standard for intraoperative neurophysiological monitoring in neurosurgery (2017 edition)[J]. National Medical Journal of China, 2018, 98(17): 1283-1293.

[48]

Zurita Perea SN, Alvarez Abut PA, Seidel K. A concise guide to D-wave monitoring during intramedullary spinal cord tumour surgery[J]. Medicina, 2024, 60(8): 1242.

[49]

Sutter M, Eggspuehler A, Muller A, et al. Multimodal intraoperative monitoring: an overview and proposal of methodology based on 1, 017 cases[J]. Eur Spine J, 2007, 16(2): 153-161.

[50]

Fehlings MG, Alvi MA, Evaniew N, et al. A clinical practice guideline for prevention, diagnosis and management of intraoperative spinal cord injury: recommendations for use of intraoperative neuromonitoring and for the use of preoperative and intraoperative protocols for patients undergoing spine surgery[J/OL]. Global Spine J, 2024, 14(): 212S-222S[2025-11-01].

[51]

Nuwer MR, Emerson RG, Galloway G, et al. Evidence-based guideline update: intraoperative spinal monitoring with somatosensory and transcranial electrical motor evoked potentials[J]. J Clin Neurophysiol, 2012, 29(1): 101-108.

[52]

Lee TK, Park EH, Lee MH. Medical ethics and artificial intelligence in neurosurgery: how should we prepare?[J/OL]. World Neurosurg, 2024, 187: e199-e209[2025-11-01].

基金资助

国家重点研发计划(2023YFC2412504┫。This study was supported by the National Key Research and Development Program of China ┣2023YFC2412504)

RIGHTS & PERMISSIONS

开放获取:本文遵循知识共享许可协议,允许第三方用户按照署名-非商业性使用-禁止演绎4.0(CC BY-NC-ND 4.0)的方式,在任何媒介以任何形式复制、传播本作品(https://creativecommons.org/licenses/by-nc-nd/4.0/)。

AI Summary AI Mindmap
PDF (542KB)

570

访问

0

被引

详细

导航
相关文章

AI思维导图

/

〈 〉