基于激光诱导击穿光谱(LIBS)与机器学习的水体低浓度铬污染等级分类
陈丁熙 , 柯川 , 何亚雄 , 温起帆 , 钱自然 , 赵栋烨
中国无机分析化学 ›› 2026, Vol. 16 ›› Issue (11) : 1990 -2001.
基于激光诱导击穿光谱(LIBS)与机器学习的水体低浓度铬污染等级分类
Classification of Low-concentration Chromium Pollution Grade in Water Based on Laser-induced Breakdown Spectroscopy (LIBS) Combined with Machine Learning
水体重金属污染是环境监测领域的重要问题,其中铬(Cr)因在电镀、冶金、皮革加工及化工等工业中的广泛应用,易通过工业废水排入自然水体,对生态系统和人体健康构成潜在威胁。为实现水体Cr污染的快速、可靠识别,本文提出了一种基于激光诱导击穿光谱(Laser-induced Breakdown Spectroscopy,LIBS)与机器学习相结合的水体Cr污染等级分类方法。首先为降低样品分布不均及等离子体波动带来的影响,本研究提出在咖啡环沉积区域进行空间扫描(Mapping)点阵取样的策略,通过10×10矩阵系统采集每个样品的100条光谱,克服样品分布不均问题。采集的光谱数据经过最大-最小归一化(Min-Max Normalization)、Savitzky-Golay(S-G)滤波及非对称最小二乘(Asymmetric Least Squares,ALS)基线校正,显著增强了特征峰强度与Cr浓度之间的线性相关性,使线性拟合决定系数R2由0.78提升至0.98。在机器学习建模环节,本研究将1 000条光谱数据随机划分为训练集、验证集和测试集,比例为8∶1∶1,分别用于模型训练、超参数调优以及最终性能评估。并选取偏最小二乘判别分析(PLS-DA)、随机森林(Random Forest,RF)、多层感知机(Multilayer Perceptron,MLP)以及支持向量机(Support Vector Machine,SVM)四种常用分类算法进行对比分析,以全面挖掘光谱特征与污染等级之间的判别关系,并评估不同算法在处理低浓度样本时的稳健性。结果显示,SVM模型在验证集和测试集上均表现出最优的分类性能,测试集整体准确率达到0.99,各污染等级样本尤其是低浓度类别样本均表现出高度稳定的判别能力。通过混淆矩阵与接收者操作特征(Receiver Operating Characteristic,ROC)曲线分析,进一步验证了SVM在类别边界判别、样本泛化能力以及多类别分类精度方面的优势。该方法不仅为工业废水中Cr污染的实时分级监测提供了可靠的技术方案,同时为LIBS技术从实验室分析向现场智能决策的转化提供了实践依据。未来可进一步结合多模态光谱数据融合,如拉曼光谱与LIBS数据联合分析,提升痕量重金属污染等级识别的精度与适用范围,从而为水环境监测和污染控制提供更高效的工具。
Heavy metal pollution in water bodies is a critical concern in environmental monitoring. Among these metals,chromium (Cr),widely used in industries such as electroplating,metallurgy,leather processing,and chemical manufacturing,can enter natural water systems through industrial wastewater,posing significant risks to ecosystems and human health. To achieve rapid and reliable identification of Cr contamination in water,this study proposed a classification method for water Cr pollution levels based on laser-induced breakdown spectroscopy (LIBS) combined with machine learning. To minimize the effects of sample inhomogeneity and plasma fluctuations,a mapping point matrix sampling strategy was employed in the coffee-ring deposition area. For each sample,100 spectra were systematically collected using a 10×10 matrix,effectively addressing sample heterogeneity. The acquired spectra were processed using min-max normalization,Savitzky-Golay (S-G) smoothing,and asymmetric least squares baseline correction. These preprocessing steps substantially enhanced the linear correlation between characteristic peak intensities and Cr concentrations,improving the coefficient of determination (R2) from 0.78 to 0.98. For machine learning modeling,1 000 spectral data points were randomly divided into training,validation,and test sets at a ratio of 8∶1∶1,used for model training,hyperparameter optimization,and performance evaluation,respectively. Four widely used classification algorithms-partial least squares discriminant analysis (PLS-DA),random forest (RF),multilayer perceptron (MLP),and support vector machine (SVM) were compared to fully explore the relationship between spectral features and pollution levels and to assess the robustness of each algorithm,particularly for low-concentration samples. The results showed that the SVM model achieved the highest classification performance on both the validation and test sets,with an overall test accuracy of 0.99. Samples from all pollution levels,especially those in the low-concentration category,demonstrated consistently stable discrimination. Analysis using the confusion matrix and receiver operating characteristic (ROC) curves further confirmed the superiority of SVM in boundary discrimination,generalization capability,and multi-class classification accuracy. This approach provided a reliable technical solution for real-time grading of Cr pollution in industrial wastewater and supports the translation of LIBS technology from laboratory analysis to on-site intelligent decision-making. Future work may integrate multi-modal spectral data,such as combining Raman spectroscopy with LIBS,to further improve the accuracy and applicability of trace heavy metal pollution level identification,thereby offering more efficient tools for water quality monitoring and pollution control.
| [1] |
Markiewicz B,Komorowicz I,Sajnóg A,et al. Chromium and its speciation in water samples by HPLC/ICP-MS-technique establishing metrological traceability:a review since 2000[J]. Talanta,2015,132:814-828. |
| [2] |
Zhang Peng,Yang Mingjie,Lan Jingjing,et al. Water quality degradation due to heavy metal contamination:health impacts and eco-friendly approaches for heavy metal remediation[J]. Toxics,2023,11(10):828. DOI:10.3390/toxics11100828. |
| [3] |
Hadi Najafabadi H,Irani M,Roshanfekr Rad L,et al. Removal of Cu2+,Pb2+ and Cr6+ from aqueous solutions using a chitosan/graphene oxide composite nanofibrous adsorbent[J]. RSC Advances,2015,5(21):16532-16539. |
| [4] |
Erdem E,Güngörmüş H,Kılınçarslan R. The investigation of some properties of cement and removal of water soluble toxic chromium(Ⅵ) ion in cement by means of different reducing agents[J]. Construction and Building Materials,2016,124:626-630. |
| [5] |
Rubio R,Sahuquillo A,Rauret G,et al. Determination of chromium in environmental and biological samples by atomic absorption spectroscopy:a review[J]. International Journal of Environmental Analytical Chemistry,1992,47(2):99-128. |
| [6] |
Rossi E,Errea M I,de Cortalezzi M M F,et al. Selective determination of Cr(Ⅵ) by on-line solid phase extraction FI-SPE-FAAS using an ion exchanger resin as sorbent:an improvement treatment of the analytical signal[J]. Microchemical Journal,2017,130:88-92. |
| [7] |
Zhang Nan,Suleiman J S,He Ming,et al. Chromium(Ⅲ)-imprinted silica gel for speciation analysis of chromium in environmental water samples with ICP-MS detection[J]. Talanta,2008,75(2):536-543. |
| [8] |
Jia Xinyan,Gong Derong,Xu Bin,et al. Development of a novel,fast,sensitive method for chromium speciation in wastewater based on an organic polymer as solid phase extraction material combined with HPLC-ICP-MS[J]. Talanta,2016,147:155-161. |
| [9] |
Frois S R,Tadeu Grassi M,de Campos M S,et al. Determination of Cr(vi) in water samples by ICP-OES after separation of Cr(iii) by montmorillonite[J]. Analytical Methods,2012,4(12):4389. DOI:10.1039/c2ay26125a. |
| [10] |
Vu H C,Dwivedi A D,Le T T,et al. Magnetite graphene oxide encapsulated in alginate beads for enhanced adsorption of Cr(Ⅵ) and As(Ⅴ) from aqueous solutions:role of crosslinking metal cations in pH control[J]. Chemical Engineering Journal,2017,307:220-229. |
| [11] |
杨增霞,宋露露,杨林. 拉曼光谱法测定废水中六价铬的含量[J]. 理化检验-化学分册,2022,58(4):485-486. |
| [12] |
Yang Zengxia,Song Lulu,Yang Lin. Determination of hexavalent chromium in waste water by Raman spectrometry[J]. Physical Testing and Chemical Analysis(Part B:Chemical Analysis),2022,58(4):485-486. |
| [13] |
Dvoynenko O,Lo S L,Chen Yijie,et al. Speciation analysis of Cr(Ⅵ) and Cr(Ⅲ) in water with surface-enhanced Raman spectroscopy[J]. ACS Omega,2021,6(3):2052-2059. |
| [14] |
Vinay B K,Bagchi S,Suranjan T R. Chromium detection in water using optical methods:a study of reagent and reagentless approaches[J]. Critical Reviews in Analytical Chemistry,2025:1-38. |
| [15] |
何亚雄,王一钦,韩晶阳, 飞秒-纳秒双脉冲激光诱导击穿光谱(LIBS)技术对合金的定量分析[J]. 中国无机分析化学,2024,14(2):214-222. |
| [16] |
He Yaxiong,Wang Yiqin,Han Jingyang,et al. Quantitative analysis of alloys by femtosecond-nanosecond dual-pulse laser-induced breakdown spectroscopy (LIBS)[J]. Chinese Journal of Inorganic Analytical Chemistry,2024,14(2):214-222. |
| [17] |
Pathak A K,Kumar R,Singh V K,et al. Assessment of LIBS for spectrochemical analysis:a review[J]. Applied Spectroscopy Reviews,2012,47(1):14-40. |
| [18] |
Casanova L,Beldjilali S A,Bilge G,et al. Evaluation of limits of detection in laser-induced breakdown spectroscopy:demonstration for food[J]. Spectrochimica Acta Part B:Atomic Spectroscopy,2023,207:106760. DOI:10.1016/j.sab.2023.106760. |
| [19] |
He Yuxin,Wen Qingfu,Ke Chao,et al. Depth-resolved and time-resolved investigations of Er2O3 ceramics after static lithium corrosion by using LIBS at different pressures[J]. Ceramics International,2023,49(22):35802-35811. |
| [20] |
Stipe C B,Miller A L,Brown J,et al. Evaluation of laser-induced breakdown spectroscopy (LIBS) for measurement of silica on filter samples of coal dust[J]. Applied Spectroscopy,2012,66(11):1286-1293. |
| [21] |
Hu Lei,Zhao Nanjing,Liu Wenqing,et al. Quantitative analysis of heavy metals in water based on LIBS with an automatic device for sample preparation[J]. Plasma Science and Technology,2015,17(8):699-703. |
| [22] |
Yu Yongli,Zhou Weidong,Qian Honggang,et al. Simultaneous determination of trace lead and chromium in water using laser-induced breakdown spectroscopy and paper substrate[J]. Plasma Science and Technology,2014,16(7):683-687. |
| [23] |
Xu Fenghua,Ma Shuxing,Tian Hongwei,et al. Detection of chromium in different valence states in water and soil using laser-induced breakdown spectroscopy combined with an ion enrichment chip[J]. Journal of Analytical Atomic Spectrometry,2023,38(7):1442-1449. |
| [24] |
Dai Xin,Song Chengbiao,Ma Shuxing,et al. Rapid determination of Cr3+ and Mn2+ in water using laser-induced breakdown spectroscopy combined with filter paper modified with gold nanoclusters[J]. Biosensors,2024,14(6):267. DOI:10.3390/bios14060267. |
| [25] |
Luo Guocheng,Zhao Changjiang,Ma Shuxing,et al. Rapid detection of total chromium in water bodies using AuAgPd trimetallic nanoparticles-modified filter paper combined with laser-induced breakdown spectroscopy[J]. Spectrochimica Acta Part B:Atomic Spectroscopy,2025,234:107336. DOI:10.1016/j.sab.2025.107336. |
| [26] |
Chen Yitao,Guo Shujie,Jiang Yufeng,et al. Direct analysis of heavy metal elements in liquid water using femtosecond laser-induced breakdown spectroscopy for high-sensitivity detection[J]. Talanta,2025,286:127512. DOI:10.1016/j.talanta.2024.127512. |
| [27] |
Chen Yitao,Jiang Yufeng,Han Junhao,et al. Sensitivity,precision,and accuracy of fs-LIBS for heavy metal detection in flowing aqueous solutions[J]. Optics Letters,2024,49(11):3106-3109. |
| [28] |
Rai N K,Rai A K. LIBS:an efficient approach for the determination of Cr in industrial wastewater[J]. Journal of Hazardous Materials,2008,150(3):835-838. |
| [29] |
Chen Yitao,Si Junrong,Wan Xingtao,et al. Sub-ppb detection limit of Cr,Pb,and Cu in water by surface-enhanced LIBS with femtosecond laser[J]. Spectrochimica Acta Part B:Atomic Spectroscopy,2024,213:106880. DOI:10.1016/j.sab.2024.106880. |
| [30] |
Sun Chao,Tian Yang,Gao Lei,et al. Machine learning allows calibration models to predict trace element concentration in soils with generalized LIBS spectra[J]. Scientific Reports,2019,9:11363. DOI:10.1038/s41598-019-47751-y. |
| [31] |
Zhang Xianglin,Zhang Feng,Kung H T,et al. Estimation of the Fe and Cu contents of the surface water in the ebinur lake basin based on LIBS and a machine learning algorithm[J]. International Journal of Environmental Research and Public Health,2018,15(11):2390. DOI:10.3390/ijerph15112390. |
| [32] |
<std>GB 3838—2002 地表水环境质量标准</std>[S].<std>GB 3838—2002 Environmental quality standards for surface water</std>[S]. |
| [33] |
<std>GB 8978—1996 污水综合排放标准</std>[S].<std>GB 8978—1996 Integrated wastewater discharge standard</std>[S]. |
| [34] |
Liu Wenjie,Sun Zongtao,Chen Jingyu,et al. Raman spectroscopy in colorectal cancer diagnostics:comparison of PCA-LDA and PLS-DA models[J]. Journal of Spectroscopy,2016,2016:1603609. DOI:10.1155/2016/1603609. |
| [35] |
Belgiu M,Drăguţ L. Random forest in remote sensing:a review of applications and future directions[J]. ISPRS Journal of Photogrammetry and Remote Sensing,2016,114:24-31. |
| [36] |
Almeida L B. Multilayer perceptrons[M]//Handbook of Neural Computation. Bristol: IOP Publishing Ltd.,2004. |
| [37] |
Pisner D A,Schnyer D M. Support vector machine[M]//Machine Learning. Amsterdam:Elsevier,2020:101-121. |
| [38] |
Suykens J A K,Vandewalle J. Least squares support vector machine classifiers[J]. Neural Processing Letters,1999,9(3):293-300. |
国家重点研发计划项目(2022YFE03020002,2022YFE03200200);四川省无人机产业创新基金资助项目(02-2025-02-0045)
/
| 〈 |
|
〉 |